Autoclave issues
All,
I want to go over some issues we've been having with autoclaves in the building:
First :
Last week we had someone drag a leaking bag of autoclave waste through a lab suite to the waste bin in the hallway. It took much effort to mop up dried tissue culture media. If you have autoclave waste, please roll the wastebin into the autoclave room to dispose of sterilized waste. Then roll it back into the hallway for disposal.
Second:
The stand alone autoclave on the third floor is not producing steam, Steris will be in tomorrow to investigate
The original autoclave on the third floor, left hand side has been running into error messages when running the waste cycle. One work around is to run antoclave waste through the wrapped cycle. This only takes and extra 10 minutes per cycle and has run through multiple cycles without issue today.
Third and probably most important:
I ran one bag through the autoclave 3 times until I discovered that the autoclave tape was not reacting (see the picture above) . Please do not buy the cheapest indicators for this vital function.
On a related note, someone is buying 45 gallon(?) autoclave bags and are filling them to the point where they will barely fit into the autoclave chamber. It will be a matter of time until one of these bags gets sucked into the evacuation port of the chamber and creates a costly and potentially dangerous situation. Please buy smaller bags to accomodate a 250 liter chamber.
Thank you all for your attention.